Data logging, analysis, and export
This DC current data logger has an accuracy of ± (0.1%+1 reading) covering its range of 0 to 20mA. Configuration settings can be managed through the DGraphPro software (available for download under the software tab). Once configured, simply press the button to begin logging immediately or schedule a delay, depending on your preferred configuration.
After the logging task is completed, connect the data logger to a PC with the DGraph Pro software installed. Run the software, and click "start" to download the data for a dlg file. When the file is opened, it can then be saved as txt, csv, xls, jpg, or bmp file format.
This software allows for efficient and comprehensive data analysis and exportation to a designated location on the user's computer.
Real-time datalogging
The real-time function is also very helpful in displaying the readings on the computer as the logger monitors the current.
Re-calibratable assures long time use in-tolerance
Each unit is shipped with a calibration report to assure accuracy from an ISO/IEC 17025:2017 conforming calibration laboratory and indicate traceability of measurements to the SI units through NIST. It can be recalibrated by DGraph Pro software with 2 points calibration by any lab with calibration capability (not recommended to be done by user) or our lab as well.
Type |
DC current data logger |
Unit |
mA |
Resolution |
0.01mA |
Measurement range |
0 to 20mA |
Accuracy |
± (0.1% + 1 reading) |
Report generation |
dlg file generated, can be saved as txt, csv, xls, jpg, bmp in the software |
Data storage capacity |
32000 |
Memory |
Non-volatile memory, no data loss when power off |
LCD display auto-off time |
30 seconds |
Sampling interval |
User configurable from 1 second to 24 hours |
High and low alarm values |
User configurable |
Alarm type |
Single time or cumulative, user configurable |
Start recording mode |
Press and hold the button until the green LED light on Start immediately or start with preset time by the software |
Stop recording mode |
Press button to stop Automatic stop when the memory is full By software when connected to computer |
Operation/storage temperature |
-30°C to 70°C (The LCD will not display normally below -10°C) |
Communication interface |
USB 2.0 |
PC Software |
DGraph Pro (Windows 7/8/Vista/10 32/64bits) website downloadable |
Power supply |
3.6V 1/2 AA lithium battery (1200mAh), user replaceable |
Battery life |
About 2 years @ 1 minute sampling interval, 25°C |
Protection Class |
IP40 |
Shell material |
ABS |
Dimensions/weight |
130mm*37mm*23mm/69g |
Certification |
CE | ROHS |
Calibration |
2 points calibration by software |
Calibration report |
Calibration report included with each unit |
Test points |
0.1, 4, 8, 12, 16 and 19.9mA on the report covering the range |