Skip to product information
1 of 6

6.3 Multi-Use Temp+RH+Dewpoint USB Data Logger w/ Display

Regular price
$108.89
Regular price
Sale price
$108.89
  • Internal RH+T sensor
  • Measurement range: -35°C to 80°C & 0.1%rh to 99.9%rh
  • Accuracy: T: ±0.3°C, RH: ±2.5%rh (20%rh to 80%rh), ±3.5%rh (else)
  • Dew point temperature data & curve output
  • PC software for configuration & data analysis & calibration
  • Storage capacity up to 32000 records
  • Non-volatile memory & No data loss when power off
  • LCD display & LED status and alarm indication 
  • Sampling interval: 2 seconds to 12 hours
  • 3.6V 1/2AA lithium battery (1200mAh), user replaceable
  • Protection Class:IP54 & Certification: CE, ROHS
  • Calibration report included with each unit
  • Test points: 20%rh, 50%rh and 80%rh @ 25°C and 25°C on the report
Sku: 36757

Shipping: 2-3 weeks

6.3 Multi-Use Temp+RH+Dewpoint USB Data Logger w/ Display
6.3 Multi-Use Temp+RH+Dewpoint USB Data Logger w/ Display

$108.89

  • Description
  • Specifications
  • Manual and Resources

Data logging, analysis, and export
This temperature and humidity data logger can be configured through the DGraph Pro software (available for download under the software tab). Once configured, simply press the button to begin logging immediately or schedule a delay, depending on your preferred configuration. 
After the logging task is completed, connect the data logger to a PC with the DGraph Pro software installed. Run the software, and click "start" to download the data for a dlg file. When the file is opened, it can then be saved as txt, csv, xls, jpg, or bmp file format.
This software allows for efficient and comprehensive data analysis and exportation to a designated location on the user's computer.
Real-time datalogging
The real-time function is also very helpful in displaying the readings on the computer as the logger monitors the temperature and humidity.
Dewpoint temperature recorded
Dewpoint temperature can be recorded and accessible in the recorded files as well as analyzed by the software
Re-calibratable assures long time use in tolerance
Each unit is shipped with a calibration report to assure accuracy from an ISO/IEC 17025:2017 conforming calibration laboratory and indicate traceability of measurements to the SI units through NIST. The temperature and humidity parameters can be recalibrated by DGraph Pro software with 2 points calibration by any lab with calibration capability (not recommended to be done by user) or our lab as well. 
Application
This temperature and humidity data logger is a good choice for a wide range of applications including labs, warehouse, production, HVAC and others.

Probe type

Internal RH+T sensor

Unit

°C, °F & %rh

Resolution

0.1°C & 0.1%rh

Measurement range

-35°C to 80°C & 0.1%rh to 99.9%rh

(Dew point temperature data & curve output)

Accuracy

T: ±0.3°C, RH: ±2.5%rh (20%rh to 80%rh), ±3.5%rh (else)

Report generation

dlg file generated, can be saved as txt, csv, xls, jpg, bmp in the software

Data storage capacity

32000 records

Memory

Non-volatile memory, no data loss when power off

LCD display auto-off time

30 seconds

Sampling interval

User configurable from 2 seconds to 12 hours

High and low alarm values

User configurable

Alarm type

Single time or cumulative, user configurable

Start recording mode

Press and hold the button until the green LED light on

Start immediately or start with preset time by the software

Stop recording mode

Press button to stop

Automatic stop when the memory is full

By software when connected to computer

Operation/storage temperature

-35°C to 80°C (The LCD will not display normally below -10°C)

Communication interface

USB 2.0

PC Software

DGraph Pro (Windows 7/8/Vista/10 32/64bits) website downloadable

Power supply

3.6V 1/2 AA lithium battery (1200mAh), user replaceable

Battery life

About 2 years @1 minute sampling interval, 25°C

Protection Class

IP54

Shell material

ABS

Dimensions/weight

120mm*37mm*23mm/67g

Certification and standards

CE | ROHS

Calibration

2 points calibration by software

Calibration report

Calibration report included with each unit

Test points

20%rh, 50%rh and 80%rh @ 25°C and 25°C on the report