Have a question?
Notified by email when this product becomes available
- Description
- Specifications
- Manual and Resources
Multi-use, Bluetooth with app, configurable
This multi-use data logger is capable of measuring temperatures with an accuracy of ±0.3°C (Optional ±0.2°C), over its range of -20°C to 70°C, with IP65 protection. The configuration settings can be managed through our PoChain software (available for download under the software tab) or our mobile app BoChain (downloadable from the App Store). Once configured, simply press the button to begin logging immediately or schedule a delay either on software settings or through the BoChain app via Bluetooth connection.
Data analysis & Data Export
Once connected to a computer with installed PoChain software, this data logger allows for efficient and comprehensive data analysis and exportation in various forms such as xls, csv, pdf, txt, and cin, to a designated location on the user's computer.
Up to 6 marker points
This data logger supports data marking up to 6 points. In the process of logging, the user can use the key to enter mark mode (MK), then press the key to mark the data.
Re-calibratable assures long time use in-tolerance
Each unit is shipped with a calibration report to assure accuracy from an ISO/IEC 17025:2017 conforming calibration laboratory and indicates traceability of measurements to the SI units through NIST. It can be recalibrated by PoChain software with up to 6 points calibration by any lab with calibration capability (not recommended to be done by user) or our lab as well.
Application
STN’s temperature logger is a good choice for a wide range of applications including transportation, storage, production, laboratory, museum, greenhouse, and others.
Probe type |
Internal sensor |
Unit |
°C, °F |
Resolution |
0.1°C |
Measurement range |
-20°C to 70°C |
Accuracy |
Basic ±0.3°C, Optional ±0.2°C |
File format |
pdf, txt, csv, xls, cin |
Data storage capacity |
39960 records |
Memory |
Non-volatile memory, no data loss when power off |
Sampling interval |
User configurable from 10s to 18 hrs |
High and low alarm values |
User configurable from -20°C to 70°C |
Start recording mode |
Press the button or by APP |
Stop recording mode |
Press the button or by APP; When connected to a computer; Automatic stop when the memory is full |
Operation temperature |
-20°C to 70°C |
Storage temperature |
-20°C to 70°C (no battery) |
Communication interface |
USB 2.0, Blue tooth 4.2 |
PC Software |
PoChain website downloadable & BoChain downloadable from APP Store |
Power supply |
CR2450 lithium battery, user replaceable |
Battery life |
1 year @ 15 minutes sampling interval, 25°C |
Protection Class |
IP65 |
Shell material |
POM |
Dimensions/weight |
φ50mm*19mm/34g |
Certification and standards |
CE | ROHS | EN12830 |
Calibration |
Max. 6 points calibration by software |
Calibration report |
Calibration report included with each unit |
Test points |
-19, 0, 25, 50 and 69°C on the report covering the range |